1
012-10116
Alkalinity ezSample Field Titrator (EZ-2340)
10–100 ppm (mg/L) CaCO
3
Safety Information
Read the MSDS before performing this test procedure. Wear safety glasses and disposable gloves.
Sample Pretreatment
If the sample is turbid, it must be filtered prior to performing this test procedure.
Test Procedure
1. Fill the sample cup to the 20 mL mark
with the sample (Figure 1).
2. Add 6 drops of A-9800 Activator
Solution (Figure 2). Stir briefly to mix
the contents of the sample cup.
NOTE: The sample should now be green. If
it is pink, total alkalinity is 0 ppm. There is
no need to continue.
3. Gently snap the tip of the ezSample
Snap Vial (ampoule) at the white ring
nearest the end of the tapered tip
(Figure 3).
NOTE: When the tip is snapped, the flexible
tubing will remain in place on the tapered
neck of the ampoule.
4. Lift the control bar and insert the
ampoule assembly into the titrator
(Figure 4).
NOTE: The rigid sample pipe will extend
approximately 1.5 inches beyond the body of
the titrator.
5. Hold the titrator with the sample pipe
in the sample and press the control bar
firmly, but briefly, to pull in a small
amount of sample (Figure 5). The
contents will turn a PINK color.
NOTE: NEVER press the control bar unless
the sample pipe is immersed in the sample.
6. With the sample pipe in the sample,
press the control bar again briefly to
allow another small amount of sample
to be drawn into the ampoule.
7. After each addition, rock the entire assembly to mix the contents of the ampoule. Watch for a color
change from PINK to
BRIGHT GREEN.
8. Repeat steps 6 and 7 until a permanent color change occurs.
9. When the color of the liquid in the ampoule changes to GREEN, remove the ampoule from the Titrator.
Hold the ampoule in a vertical position and read the scale opposite the liquid level (Figure 6).
Figure 5
Figure 2
Figure 1
Figure 3
Figure 4
Figure 6
Vista de pagina 0
1 2

Indice de contenidos

Pagina 1 - Test Procedure

1 012-10116 Alkalinity ezSample™ Field Titrator (EZ-2340) 10–100 ppm (mg/L) CaCO3 Safety Information Read the MSDS before performing this test pr

Pagina 2 - Test Method Description

2 012-10116 Test Method Description The alkalinity ezSample test method employs an acid titrant and a mixed pH indicator.1–3 Results are expresse